The EnviteC Medical Oxygen Sensors are intended as oxygen sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures in the following applications: Sensing device for oxygen in medical ventilators, anesthesia machines, incubators, and control device of oxygen concentrators.
Compatibility:
- Hamilton HM-01, Veolar, Amadeus, Arabella, Novacare, Hudson 5568, C1, C-01, Oxequip Oxitron 3250, Vickers incubator, Hamilton HM-11, SS01 (SS-01),MAX19, PSR-11-33-1, 10-103-00, PT01, Teledyne T1 A-37016
Use the advantages:
- Compliant with European MDD (CE certification)
- Meets ISO 80601-2-55
- Designed and manufactured according to EN ISO 13485
- Accurate and reliable fast response
- Resistant to N2OTechnical support
- Made in Germany
- FDA cleared
Technical Specifications OOM202-2
Measurement range 0 % … 100 % oxygen (at atmospheric pressure)
Nominal sensor lifetime ≥ 1 000 000 % volume oxygen hours
Output in ambient air 9 mV … 13 mV
Electrical interface Flying leads with 3 pin female connector (Molex® 22-01-2037)
Accuracy meets ISO 80601-2-55 requirements
Repeatability < 1 % volume O2 at constant temperature and pressure
Linearity error < 3 % relative
Response time < 12 s to 90 % of final value
Zero offset voltage < 200 µV in 100 % nitrogen, applied for 5 min
Cross interference meets ISO 80601-2-55 requirements
Influence of humidity -0.03 % rel. per % RH at 25 °C
Pressure range 0.6 bar … 2 bar (ppO2 0 … 1250 mbar O2)
Influence of pressure proportional to change in oxygen partial pressure
Influence of mechanical shock < 1 % relative after a fall from 1 m
Operating temperature 0 °C … +50 °C
Temperature compensation built-in NTC compensation
Effect of temperature compensation (steady state)
between +25 °C and +40 °C: 3 % relative error
between 0 °C and +50 °C: 8 % relative error
Operating humidity 0 % … 99 % RH non-condensing
Long term output drift < 1 % volume oxygen per month
typically < -15 % relative over lifetime
Storage temperature -20 °C … +50 °C
Recommended storage +5 °C … +15 °C
Recommended load ≥ 10 kOhms
Warm-up time < 30 minutes, after replacement of sensor
Weight approximately 28 grams