In Stock

Envitec OOM202 Medical Ventilator Oxygen Sensor

 27,500

Envitec Honeywell OOM202 Medical Ventilator Oxygen Sensor

Made in Germany

The EnviteC Medical Oxygen Sensors are intended as oxygen sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures in the following applications: Sensing device for oxygen in medical ventilators, anesthesia machines, incubators, and control device of oxygen concentrators.

EnviteC by Honeywell

Envitec OOM202 is compatible with Arduino

Oxygen sensor OOM202 for Ventilator OOM202 O2 sensor O2 cell ENVITEC Germany Oxygen battery 00M202

Features:
1.Replacement part for Oxygen sensor OOM202
2. Comes in Sealed Tin Can

Compare
SKU: OOM202-NENOPK Categories: ,

Description

Product Description

OOM 202 For SIEMENS, Hamilton, Medisys, Agva, Puritan Bennette Oxygen Sensor.
The EnviteC Medical Oxygen Sensors are intended as oxygen sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures in the following applications: Sensing device for oxygen in medical ventilators, anesthesia machines, incubators, and control device of oxygen concentrators.

Compatibility:

  • Hamilton HM-01, Veolar, Amadeus, Arabella, Novacare, Hudson 5568, C1, C-01, Oxequip Oxitron 3250, Vickers incubator, Hamilton HM-11, SS01 (SS-01),MAX19, PSR-11-33-1, 10-103-00, PT01, Teledyne T1 A-37016

Use the advantages:

  • Compliant with European MDD (CE certification)
  • Meets ISO 80601-2-55
  • Designed and manufactured according to EN ISO 13485
  • Accurate and reliable fast response
  • Resistant to N2OTechnical support
  • Made in Germany
  • FDA cleared

Technical Specifications OOM202-2
Measurement range 0 % … 100 % oxygen (at atmospheric pressure)
Nominal sensor lifetime ≥ 1 000 000 % volume oxygen hours
Output in ambient air 9 mV … 13 mV
Electrical interface Flying leads with 3 pin female connector (Molex® 22-01-2037)
Accuracy meets ISO 80601-2-55 requirements
Repeatability < 1 % volume O2 at constant temperature and pressure
Linearity error < 3 % relative
Response time < 12 s to 90 % of final value
Zero offset voltage < 200 µV in 100 % nitrogen, applied for 5 min
Cross interference meets ISO 80601-2-55 requirements
Influence of humidity -0.03 % rel. per % RH at 25 °C
Pressure range 0.6 bar … 2 bar (ppO2 0 … 1250 mbar O2)
Influence of pressure proportional to change in oxygen partial pressure
Influence of mechanical shock < 1 % relative after a fall from 1 m
Operating temperature 0 °C … +50 °C
Temperature compensation built-in NTC compensation
Effect of temperature compensation (steady state)
between +25 °C and +40 °C: 3 % relative error
between 0 °C and +50 °C: 8 % relative error
Operating humidity 0 % … 99 % RH non-condensing
Long term output drift < 1 % volume oxygen per month
typically < -15 % relative over lifetime
Storage temperature -20 °C … +50 °C
Recommended storage +5 °C … +15 °C
Recommended load ≥ 10 kOhms
Warm-up time < 30 minutes, after replacement of sensor
Weight approximately 28 grams
Part number 01-00-0068